Home > Producten > Analyzers > 1231 & 1232 In-Situ Oxygen Probes

1231 Oxygen probe without filters

1231 & 1232 In-Situ Oxygen Probes

The 1231 & 1232 In-Situ Oxygen Probes from Novatech Controls, available through Novoxys, deliver unparalleled performance for oxygen measurement in demanding industrial processes. Featuring the world’s most robust zirconia sensors, these probes provide rapid response times and minimal drift.

Choose the heated 1231 for precise flue gas analysis up to 900°C, or the 1232 for extreme temperatures up to 1400°C in industrial boilers, furnaces, and incinerators. Both models ensure maximum measurement accuracy, optimize combustion efficiency, and enhance safety by helping prevent dangerous fuel-rich conditions. A reliable and durable solution for optimal process control.

Main features of the Novatech 1231 & 1232 In-Situ Oxygen Probes:

  1. Ultra-Robust Zirconia Sensors: Built with high-durability zirconia technology, offering exceptional resistance to thermal shock and harsh industrial environments with minimal drift.

  2. Extreme Temperature Capabilities: The heated 1231 operates seamlessly up to 900°C, while the unheated 1232 is engineered for high-temperature applications up to 1400°C.

  3. Rapid Response Time: Delivers fast, real-time oxygen measurements to enable immediate adjustments to the combustion process.

  4. In-Situ Measurement: Direct flue gas analysis within the duct or stack eliminates the need for complex, high-maintenance sample extraction systems.

  5. Combustion Efficiency & Safety Optimization: Ensures precise air-to-fuel ratio control, reducing fuel consumption and emissions while preventing hazardous fuel-rich conditions.

Model 1231 Probe

CharactisticSpecification
ApplicationCombustion flue gasses below 900°C
Temperature range0°C to 900°C
Probe insertion lengthStandard lengths: 250, 350, 500, 750, 1000, 1500mm
Process connection1.5” BSPT
Outer sheath:32mm OD 316 s/steel, 42mm with optional filter
Electrical connectionWeather-proof plug-in connector or optional screw terminals. The plug connector can be supplied with the cable
Internal thermocoupleType ‘K’
inute Ref. air connection1/4” tube
Cal. check gas connection1/8” NPT female
WeightApprox. 2kg + 165g per 100mm insertion length

Model 1232 Probe

CharactisticSpecification
ApplicationCombustion flue gasses above 700°C (1290°F) with no contaminants; Eg natural gas, light oils
Temperature range253MA stainless steel 700°C to 1100°C
Temperature rangeAlumina ceramic 700°C to 1400°C
Probe insertion lengthStandard lengths: 500, 750, 1000mm (20”, 30”, 40”)
Process connection3/4” BSPT
Outer sheath:19mm OD 253MA stainless steel or alumina ceramic
Head temperature-25°C to 150°C ( -15°F to 300°F)
Internal thermocoupleTypically Type ‘R’. Also available with no TC, Type ‘S’ or ‘K’
inute Ref. air connectionIntegral air line through connector or 1.4” tube
Cal. check gas connection1/8” NPT female
WeightApprox. 1kg + 100g per 100mm insertion length

Downloads

Information Request